Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK CSG110 Product Manual
Drawing Number: EM1000-E - 10 - Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK Appendix A (Noise levels) Current Outp
Drawing Number: EM1000-E - 11 - Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK Appendix B (Specifications) Electrical Spe
- 2 - Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK Table of Contents Default Settings ...
- 3 - Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK Default Settings Input Range: 0 to +/-2 mV/V Excitation Voltag
- 4 - Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK Standard Span & Zero Adjustment Once all of the connections are com
- 5 - Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK Shunt Readings Shunt resistors simulate a load on the load cell; thus,
- 6 - Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK shuntexcitationpolaritybandwidthgain SW1SW2SW4SW5current settingSW8SW6c
- 7 - Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK Ex. If you are using your CSG110 with a tension and compression load ce
- 8 - Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 1 UP 0.5 5 UP 2.5DOWN N/ A DOWN N/ A2 UP 1 6 UP 3DOWN N/ A DOWN N/ A3 U
- 9 - Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK Advanced Span and Zero Adjustment Adjusting the Zero At times, when usi