Millivolt per Volt Calibration Menu Structure
To enter the MilliVolt per Volt Calibration Menu, press and hold the and buttons for 3 seconds.
10 Thomas, Irvine, CA 92618 USA T (949)-465-0900 F (949)-465-0905 [email protected] www.futek.com/support/IHH200
IHH200 Manual 2.5 - Page 15
Parameter
Set-Up Information
Press To skip to next menu item
Press To change the 5mV/V gain.
Here the factory gain calibration can be changed to a measured value (see Milli-Volt
Calibration Procedure - back end of manual).
Once the derived value has been entered press to confirm.
Press to skip to next menu item
Press to change the 5mV/V offset.
Here the factory offset value can be changed to a measured value (see Milli-Volt
Calibration Procedure - back end of manual).
Once the derived value has been entered press to confirm.
THIS CAN ONLY BE SET WHEN USING 50mV/V RANGE
Press To skip to next menu item
Press To change the 50mV/V gain.
Here the factory gain calibration can be changed to a measured value (see Milli-Volt
Calibration Procedure - back end of manual).
Once the derived value has been entered press to confirm.
THIS CAN ONLY BE SET WHEN USING 50mV/V RANGE
Press to skip to next menu item
Press to change the 5mV/V offset.
Here the factory offset value can be changed to a measured value (see Milli-Volt
Calibration Procedure - back end of manual).
Once the derived value has been entered press to confirm.
5.0 gAIn
5.0 OFFS
50 gAIN
50 OFFS