Calibration Menu
To enter the CALIBRATION MENU, press and hold the and buttons for 3 seconds.
Press To skip to next menu item
Press To change sensor input sensitivity
This allows the calibration engineer to change the sensitivity range of the IHH200 when
connecting to sensors with a sensitivity of greater than 5mV/V. The IHH200 is factory set
for 5mV/V. To ensure the unit is set to 5mV/V press
To select 50mV/V you need to power down the unit and access the internal circuit board.
Move link LK1 and place it onto JP1. Power on the IHH200 and return to this point of the
calibration menu. You will notice that the menu parameter has changed to SEnS 50.0;
press to change the sensitivity to 50mV/V and move on to the next parameter.
Press to skip to next menu item
Press to the set the display resolution
This parameter sets the decimal point position for the display and the resolution, i.e. a
value of 000.005 would display the reading to 3 decimal places and the readings will
change in steps of 0.005.
The decimal point position is moved one place to the right each time you press the
and together.
Any value can be entered for the resolution, using the and arrows to select a
digit and the and arrows to increment or decrement the digits. Press to
accept the value and move onto the next parameter.
To save the settings and move on to the next parameter press
THIS MENU IS DISABLED WHEN TEDS IS ENABLED
Press to skip to next menu item.
Press to the enter the calibration routine
If you have chosen to enter the calibration routine you will be prompted whether you want
to select LiVE. If you do not press , press . You will then be prompted to select
either of the other calibration methods, which in order, are tAbLE and CAL VAL. To select
any of the calibration methods press . Otherwise press
For more detailed calibration information, please refer to the calibration section of the
manual.